From the Department of Obstetrics and Gynecology, Derby City General Hospital, Derby, United Kingdom.
Address reprint requests to: Richard Hayman, MRCOG, DM, City Hospital, Department of Obstetrics and Gynecology, Nottingham, NG5 1PB, United Kingdom; E-mail: richard.hayman{at}nottingham.ac.uk.
| ABSTRACT |
|---|
|
|
|---|
METHODS: Between December 1999 and September 2000, a prospective audit was undertaken of all vacuum deliveries performed at Derby City General Hospital.
RESULTS: In this period, 3296 deliveries occurred, of which 317 (9.6%) were by vacuum. Of these, data were collected on 300 (94.6%), 78 deliveries with the hand-held vacuum and 222 with standard vacuum. There were no differences in the demographic profiles, indication, gestational age at delivery, or birth weights between the two groups (P > .05 in all instances). In all types of delivery, nonrotational and rotational, the hand-held vacuum performed comparably to its contemporaries with no increase in delivery "failures" being noted. There were no differences in the extent or frequency of maternal injuries between the instruments, and other than transient scalp abrasions, there were no significant fetal injuries.
CONCLUSION: The hand-held vacuum delivery system is a functionally effective addition to the practitioners "armory," providing an alternative to the standard metal and silastic cups.
Operative vaginal delivery has been clearly identified as a major risk factor for fetal morbidity and mortality as well as early and late maternal morbidity (including fecal incontinence). The Royal College of Obstetricians and Gynecologists issued clinical guidelines regarding the use of these instruments in assisting vaginal delivery, examining their relative merits, indications for use, and their associated complications. They concluded by stating that: "Obstetricians should be competent, and confident, in the use of both vacuum and forceps,"1,2 but "in view of the reduction of maternal injuries the vacuum should be considered to be the instrument of first choice."3
Consequent upon this article, interest in the practice of vacuum-assisted vaginal deliveries increased. However, serious neonatal and maternal complications with the incorrect or inappropriate use of a vacuum device reinforced the need for appropriate training in the correct use of this equipment. Attempts have been made to identify and address the predisposing factors associated with the adverse outcomes,4 and several novel changes have been made to the basic instrument design to facilitate their use and reduce the rate of complications.
The basic premise of any vacuum device is that a suction cup, of silastic or metal construction, is connected, via tubing, to a vacuum source. Either directly, though the tubing, or via a connecting "chain," traction can then be applied to the presenting part, the vertex, to expedite delivery.
For successful use of the vacuum, determination of the flexion point is vital. This is located, in an average term infant, on the sagital suture 3 cm anterior to the posterior fontanelle, and thus 6 cm posterior to the anterior fontanelle. The center of the cup should be placed directly over this, as failure to adequately position the cup can lead to a progressive deflexion of the fetal head during traction, and failure to deliver the baby.5 However, not all vacuum cup designs allow easy positioning over the flexion point, especially when the fetal head adopts an occipito-posterior or lateral position. This problem is frequently compounded by the presence of asynclitism or deflexion.
The Kiwi OmniCup (Clinical Innovations, Murray, UT) (Figure 1
) is a vacuum extraction device that incorporates an integral hand-held pump (PalmPump; Clinical Innovations), making it suitable for single-person use with the addition of a "posterior cup" design that makes it usable for all positions of the vertex.6 In addition, with the theoretical potential for the transmission of pathogenic organisms between individuals, even with equipment that has undergone surgical standard sterilization, single-use devices should replace the standard reusable equipment.
|
| MATERIALS AND METHODS |
|---|
|
|
|---|
After a brief period of familiarization, the OmniCup was made available to all labor ward practitioners for use alongside the "standard equipment" when a valid indication for vacuum-assisted delivery was present. No specific additional training was given to the practitioners using the Kiwi OmniCup as they had been trained in the use of other types of vacuum cups/devices and were already using them on a daily basis.
The type of cup chosen by the practitioner (Silc, Malmstrom, Bird, or the OmniCup) was based on the obstetric scenario and the clinicians level of experience and training. At no stage was the practitioner instructed to use the OmniCup vacuum in preference to any other, and in all cases the procedure was performed according to standard hospital protocol for vacuum extraction.
The ethics committee of Derby City General Hospital determined that as the equipment had previously been shown to be "safe,"6 special consent to use a specific cup did not have to be obtained from the laboring women for inclusion in this prospective audit.
After the delivery, all the relevant obstetric details were recorded onto a standardized prospective clinical questionnaire. Information regarding the performance of each vacuum, its placement with respect to the vertex, and any complications experienced by the mother and the newborn infants was also collected. Infants were also assessed at 2448 hours postpartum, and any specific fetal problems noted at that time were reported to the general practitioner/visiting midwife for further monitoring or referral as appropriate.
The primary outcome measures for each instrument used were delivery failure/success rates and fetal and maternal complications.
The successful background rate of vacuum deliveries was observed to be 10% (standard deviation 5%) in the period before the onset of the audit project. It was calculated that a sample size of 130 women (65 in each arm) would be required to observe a difference of 25% between the two cohorts regarding failure of vacuum deliveries with an error of 5% and a power of 80%.
| RESULTS |
|---|
|
|
|---|
Deliveries with Silc, Malmstrom, or Bird cups are included in the "standard vacuum" group. Deliveries by the OmniCup vacuum are analyzed separately.
The indications and relevant obstetric circumstances relating to the 300 vacuum deliveries are shown in Table 1
. There were no significant differences in the parity, age of the patients, or gestation at delivery (P > .05, Mann-Whitney U test), in the type of labor (P = .36,
2 test), analgesic requirements (P = .59,
2 test), or median birth weights (P = .54, Mann-Whitney U test).
|
2 test).
In the majority of cases with both instrument "groups," the position of the fetal head at the time of cup application was occipito-anterior, although there was a slight but not significant increase in the number of rotational deliveries (occipito-lateral and posterior) performed with the OmniCup (P = .105) (Table 1
).
The status of the operators who performed the deliveries is shown in Table 2
. The majority of deliveries were performed by residents in their fourth to tenth years of training, the remainder being by residents in years 13 (Table 2
).
|
2 test). There was a slight, but nonsignificant, increase in the number of deliveries with the OmniCup, which had to be converted to forceps deliveries (P = .12,
2 test). Subanalysis of the data revealed that this "failure rate" was similar to that observed for the Bird design of metal cup (P = .13,
2 test). However, the number of delivery failures (defined as failure of vaginal delivery with resort to delivery by cesarean) with the standard equipment (2.3%) exceeded those with the OmniCup vacuum (1.3%). This, too, did not reach statistical significance (P = .29,
2 test).
|
Fetal complications were also infrequent and are detailed in Table 4
. There were no differences in the Apgar scores at 1 and 5 minutes between the groups (P = .18,
2 test). The only significant finding was an increase in the incidence of superficial scalp abrasions (minor graze or linear laceration underlying the vacuum cup margins) in the cohort delivered with the OmniCup vacuum (14.1% versus 4.5% in the standard group, P = .006,
2 test). All of these abrasions resolved spontaneously with no scarring noticeable at 6 weeks postpartum. It was interesting to observe that incidence of scalp abrasions was the same in the OmniCup cohort as in those infants delivered by rotational metal cup vacuum devices (14.1% versus 11.1%, P = .39,
2 test). All the cephalhematomas that developed were transient and were not linked to the development of clinically significant jaundice in any instance.
|
| DISCUSSION |
|---|
|
|
|---|
It was, therefore, important to observe that in no instance, with any of the equipment used during the course of our audit, did any significant complications occur to either the mother or infant. Although the incidence of scalp abrasions in those infants delivered by the OmniCup was higher than in those delivered by the alternative standard equipment, it was the same as those in the infants delivered by rotational metal cup vacuum devices. This would suggest that it is the process of rotation that is implicated in the generation of the trauma rather than the design of the device itself. However, in all cases, the degree of injury sustained was minor and had resolved completely by 6 weeks postpartum.
The slight increase in the number of forceps deliveries after use of the OmniCup when compared with the "standard" vacuum may be explained by examining the indications for delivery. In such cases, it was observed that a significant number of vacuums were being employed to rotate the presenting part to a more favorable position (occipito-anterior) before procuring delivery with forceps. In fact, subanalysis of the data confirms this to be the case, with metal cup vacuum of the Bird design having a similar incidence of subsequent forceps deliveries when compared with the OmniCup.
A limitation of this study concerns the nonrandomization of patients to the different delivery device cohorts. This has the potential to introduce a degree of selection bias into data sets, with allocation into each group being on the basis of ease of application/operator confidence. A further randomized study will need to specifically address these issues, as this initial audit cannot exclude this potential problem.
Data from previous studies have shown that procuring a delivery is highly dependent upon the correct placement of a vacuum over the flexion point.5,6,9 The OmniCup design used in this study facilitates placement over the flexion point akin to the Bird cup currently in use for occipito-posterior positions. Unlike its predecessors, however, the inclusion of a hand-pump vacuum generation system decreases both the "clumsiness" of the additional equipment required to generate the suction and the extra personnel needed to operate the pump. In addition, the flexibility of the traction axis device engendered the OmniCup to the practitioners, enabling its use in all clinical vertex presentations, both "lift out" and rotational.
This study suggests that the Kiwi OmniCup is an effective instrument for assisting a vaginal delivery. As with all instruments used in delivery, however, the device must be used appropriately and correctly, with strict adherence to the recommended safeguard for vacuum-assisted vaginal delivery.6
| Footnotes |
|---|
Received March 13, 2002. Received in revised form June 10, 2002. Accepted June 27, 2002.
| REFERENCES |
|---|
|
|
|---|
2. Gei AF, Belfort MA. Forceps-assisted vaginal delivery. Obstet Gynecol Clin N Am 1999;26:34570.[Medline]
3. Johanson RB. Instrumental vaginal delivery. London, United Kingdom: Royal College of Obstetricians and Gynaecologists, 1999.
4. Vacca A. The trouble with vacuum extraction. Curr Obstet Gynecol 1999;9:4155.
5. Bird GC. The importance of flexion in vacuum extraction delivery. Br J Obstet Gynaecol 1976;83:194200.[Medline]
6. Vacca A. Operative vaginal delivery: Clinical appraisal of a new vacuum extraction device. Aust N Z J Obstet Gynaecol 2001;41:15660.[Medline]
7. Clinical Effectiveness Support Unit. The use and interpretation of cardiotocography in intrapartum fetal surveillance. London, United Kingdom: Royal College of Obstetricians and Gynaecologists, 2001.
8. American College of Obstetricians and Gynecologists. Operative vaginal delivery. Technical bulletin no. 152. Int J Obstet Gynaecol 1992;38:5560.
9. Bird CG. The use of the vacuum extractor. Clin Obstet Gynecol 1982;9:64161.